We sense good chemistry...

Detailed information
Title Development of a Solid Phase Extraction technique (SPE) for 257 compounds suitable for both online SPE tandem Q-TOF analysis and also for advanced toxicity bio-monitoring
Summary An extensive study has been performed in order to identify the most efficient combination of SPE material and pH for retrieving the maximum of the tested polar compounds The development of this method with a broad applicability to such a great range of compounds improves the potential of its integration to online toxicity biomonitors. Consequently, lowering the detection limits of biomonitors will result in receiving better early warning alarms.
Category 3. Symbiosis User Application

These are applications that have been made by Symbiosis customers. They are validated and have been published.
 
Matrix drinking water, potable water
Market Environmental
Technique online SPE
Cartridge OASIS HLB, HySphere resin-SH
Detection Q-TOF (Agilent)
Compounds 257 trace contaminants such as pesticides, herbicides and pharmaceuticals
Compound Group trace contaminants. pesticides, herbicides and pharmaceuticals
Sample_volume not provided
 
Contact Varvara Kokkali
Account Vitens, Snekertrekweg 61, 8912AA Leeuwarden, The Netherlands
Published Poster presented by Vitens in 2013
Date 25.07.2013
  Read PDF document
 
Remarks Tested SPE cartridges: Oasis® HLB HySphere Resin-SH HySphere Resin-GP HySphere C18-HD HySphere C8-EC-SE HySphere MM Anion HySphere MM Cation HySphere C2-SE HySphere CN-SE PLRP-s XB 55 XB 96 Isolute® NH2

Users
Status Published