Detailed information | |
Title | Simultaneous, Sensitive LC/MS Analysis of Plasma Metanephrines |
Summary | in this study, a clinical research method has been developed for the simultaneous measurement of three plasma biogenic amines. The method utilizes LC/MS and an online SPE system. This combination of SPE sample preparation coupled with the analyticla power of LC/MS is able to deliver an efficient assay for these important metanephrines |
Category | 2. Symbiosis Application Notes These are application notes that have been made in our application lab. They were based on customer demo's and customer method development requests. They include sorbent screening, a calibration curve and are optimized for carry over. They are not always (completely) validated |
Matrix | Plasma |
Market | OSM Clinical |
Technique | online SPE |
Cartridge | Waters MassTrak WCX OSM cartridge |
Detection | Waters XEVO TQ-S |
Compounds | Metanephrine, normetanephrine and 3-methoxytyramine, 3-MT |
Compound Group | Metanephrines |
Sample_volume | |
Contact | Jeanne Adaway and Brian Keevil |
Account | University Hospital South Manchester, UK |
Published | Poster, Technical Brief |
Date | 04.08.2015 |
Read PDF document | |
Remarks | |
Users | |
Status | published |