Detailed information | |
Title | Monitoring Opioid Use for Pain Management Using a Fully Automated DBS-SPE-LC/MS/MS Approach |
Summary | An automated DBS-SPE-LC/MS/MS method for the quantitation of four representative opioids in whole blood was developed and validated over the concentration range of 1 to 500 ng/mL blood, allowing detection of subtherapeutic, therapeutic, and toxic levels. |
Category | 1. Prospekt 1 Application information These applications were mostly done on Prospekt1 and can offer a good starting point when executed on Symbiosis. Care should be taken on differences in volume (SDU vs. HPD). Also, most information is based on other detectors than MS |
Matrix | blood |
Market | Clinical |
Technique | DBS |
Cartridge | HySphere C18 HD, 7 μm, 2 x 10 mm (Spark-Holland) |
Detection | LCMS-8050 mass spectrometer (Shimadzu) |
Compounds | morphine, codeine, oxycodone, hydrocodone |
Compound Group | opioids, drugs |
Sample_volume | 10 µL |
Contact | Ruth Verplaetse and Jack Henion |
Account | Quintiles, 19 Brown Road, Ithaca, New York |
Published | poster |
Date | 10.12.2019 |
Read PDF document | |
Remarks | |
Users | |
Status | poster |